Service Manual
Test Trigger Sensitivity
Steps
1Press the button “DEFAULT/SETUP”, and set the VOLT/DIV to 500mv/div, set
SEC/DIV to 25ns/div.
2Connect the oscilloscope channel 1 to the standard sinewave generator
3Set the standard sinewave generator frequency to 10 MHz
3Set the generator output level so the Vpp measurement is about 500 mV.
4Press the button “set to 50%”. Adjust LEVEL as necessary and then check that
triggering is stable.
5、Set the generator frequency to:
25MHz if you are testing a SDS1022D
40MHz if you are testing a SDS1042D
60MHz if you are testing a SDS1062D
100MHz if you are testing a SDS1102D
150MHz if you are testing a SDS1152D
200MHz if you are testing a SDS1202D
6Set the generator output level so the Vpp measurement is about 750 mV.
7Press the button “set to 50%”. Adjust LEVEL as necessary and then check that
triggering is stable.
8Test channel 2 with the same method
Test Noise
Steps 1 Set the both channels VOLT/DIV to 2mV/divSEC/DIV to 2.5µS/div
2 Set two measure parameters as “CH1,Vpp” and “CH2, Vpp”
3 Check the Vpp measurement is 1.2mV other VOLT/DIV is 0.6 DIV
4 Change the VOLT/DIV, and check the Vpp until all the VOLT/DIV of both channel
have been tested.